Description |
Module for measuring the hydrostatic pressure relatively to
the atmospheric air pressure of the site and the water
temperature. The hydrostatic pressure is a measurement for the
height of the water column, weighing on the sensor. |
A/D converter |
16 Bit integrating |
Interface |
AquiTronic sensor bus interface |
Measuring principle |
Enclosed piezo-resistive pressure measuring cell with
compensation of temperature |
Material |
Titan |
Accuracy |
0.1% of final value |
Resolution |
0.05% of measuring range |
Overload |
up to max. 4-fold measuring range, indication of measuring
value up to 1.5 fold measuring range |
Compensated
temperature range |
0 to +50 °C |
Available measuring range |
100,200,350,500,700,1000,2000,3500 and 5000 kPa (100 kPa = 1
bar corresponding to approx. 10m water column) |
|
|
principle of measuring |
PT1000 |
Measuring range |
0 to +50 °C |
Resolution |
0.025 °C |
|
|
Housing material |
V4A high-grade steel, POMŽ |
Sealing |
VitonŽ |
Dimensions |
Total length: 120mm
Length in probe: 105mm
Diameter: 32mm |
Weight |
250g |
Operating temperature |
-20 to +50 °C |
|
|